Gw Instek GDS-2302A 2-channel oscilloscope, Digital Storage oscilloscope, GDS-2302A Datenbogen

Produktcode
GDS-2302A
Seite von 261
 
 
SAVE/RECALL 
187 
 
Ref 1~4 
Reference waveforms stored in the 
internal memory, separate from W1 ~ 
W20. Reference waveforms (Ref 1 ~ 4) 
can be displayed directly onto the 
display with amplitude and frequency 
information. Ref 1~4 are useful for 
reference purposes. Other waveforms ( 
LSF and W1~20) must be recalled to 
R1~4 before being displayed. 
Contents: 
Waveform Data 
The waveform data can be used for detailed 
analysis. It consists of the horizontal and vertical 
data used by the waveform.  
Spreadsheet File Format 
 
Format 
DSxxxx.csv (Comma-separated values format, can 
be opened in spreadsheet applications such as 
Microsoft Excel).  
CSV-formatted files can be stored in either a short-
memory format or a long-memory format: Detail 
CSV, Fast CSV, LM Detail CSV and LM Fast CSV. 
Detail CSV will record both the horizontal and 
vertical sample points of the waveform. All the 
points are recorded in scientific notation for analog 
data. CSV data will only record 5000 points of data.  
Fast CSV will only record the vertical amplitude of 
the sample points. Fast CSV also contains data that 
enables the horizontal data points to be 
reconstructed, such as trigger position, etc. Fast 
CSV will only record 5000 points of data. Data is 
recorded as integers. 
LM Detail CSV is similar to CSV but covers the 
entire long-memory (See page 109 for information 
on the record length). All points are recorded in 
scientific notation for analog data. 
LM Fast CSV is similar to Fast CSV but covers the