DS1000Z Quick Guide 
Probe Compensation 
 
When the probes are used for the first time, you should 
compensate the probes to match the input channels of the 
oscilloscope. Non-compensated or poorly compensated probes 
may cause measurement inaccuracy and error. The probe 
compensation procedures are as follows: 
 
1.  Perform step 1, 2, 3 and 4 of “Function Inspection”
2.  Check the displayed waveforms and compare them with the 
following figures. 
 
 
 
 
 
Figure 8 Probe Compensation 
3.  Use a nonmetallic driver to adjust the low-frequency 
compensation adjustment hole on the probe until the 
displayed waveform is as the “Correctly compensated” in the 
figure above. 
 
 
 
 
 
 
 
 