Rigol DS1104Z 4-channel oscilloscope, Digital Storage oscilloscope, DS1104Z Benutzerhandbuch

Produktcode
DS1104Z
Seite von 264
RIGOL 
DS1000Z User’s Guide 
6-28 
Statistic Function 
 
Make statistic and display the current, average, minimum (or standard deviation) and 
maximum (or count) values of at most 5 measurement items that are turned on last.   
Press Measure  Statistic to turn the statistic function on or off. When the 
statistic function is enabled, press Stat.Sel to select “Extremum” or “Difference” 
measurement. When “Extremum” is selected, minimum and maximum values are 
displayed. When “Difference” is selected, standard deviation and count values are 
displayed. 
 
 
Extremum Measurement 
 
Difference Measurement 
 
Press Measure  Reset Stat to clear the history data and make statistic again.