4.  MISO Setting 
Press MISO to select any channel (CH1-CH4) as the MISO data channel. When 
“OFF” is selected, this data line is not set. 
 
5.  MOSI Setting 
Press MOSI to select any channel (CH1-CH4) as the MOSI data   
 
6.  Edge Setting 
Press Edge to sample the MISO and MOSI at the rising edge or falling edge of 
CLK. 
 
7.  Polarity setting 
Press Polarity to set the polarity of the SDA data line to 
  (the high level is 1) 
or 
  (the low level is 1). 
 
8.  Width Setting 
Press Width to set the bit width of each frame. The range available is from 8 to 
32. 
 
9.  Endian Setting 
Press Endian to select “LSB” or “MSB” and the default is “MSB”. 
 
10.  Display-related Setting 
  Press Format to set the display format of the bus to Hex, Decimal, Binary, 
ASCII or LINE. Note that LINE format is displayed the acyual value of the 
bus in a binary number form, and the order is consistent with the bus 
transmission sequence. This format is only valid for the serial bus, because 
the serial bus has the difference of bit sequence of LSB and MSB; if bit 
sequence of the bus is selected MSB, the LINE format is the same to the 
binary format. 
  Press Pos and use 
  to adjust the vertical display position of the bus. 
 
11.  Decoding Configuration 
Press Decoding1 Configure to turn the decoding configuration submenu on. 
  Press Label to turn the label display function on or off. When it is turned on, 
the bus label will be displayed at the upper left of the bus (when the bus 
display is turned on).