DS1000Z User’s Guide
 
 
1-13  
 
signal and the result is equal to the current capture rate. 
Pass/Fail: there will be output a negative pulse signal when failed waveforms 
are detected by the oscilloscope. Connect this signal to other control systems to 
view the test result conveniently. There will be output a 3.3 V CMOS high level 
when no failed waveform is detected by the oscilloscope. 
 
5.  Signal Output 
When the output of Source1 and Source2 are enabled, the signal currently set 
can be output to the analog input terminal of the oscilloscope or external 
devices connected to them through the [Source1] and [Source2] connectors 
at the rear panel. 
 
6.  Lock Hole 
You can lock the instrument to a fixed location using the security lock (please 
buy it yourself) via the lock hole. 
 
7.  Fuse 
If a new fuse is required, please use the specified fuse (250V, T2A). 
a)  Turn off the instrument and remove the power cord. 
b)  Insert a straight screwdriver into the slot at the power socket and prize out 
the fuse seat gently. 
c)  Take out the fuse and replace it with specified fuse, and then install the fuse 
seat to the original position. 
 
8.  AC Power Socket 
AC power input terminal. The power requirements of this oscilloscope are 
100-240 V, 45-440 Hz. Use the power cord provided with the accessories to 
connect the instrument to AC power. Then, you can press the power key at the 
front panel to start the instrument.