DS1000Z User’s Guide 
4-2 
Acquisition Mode   
 
The acquisition mode is used to control how to generate waveform points from 
sample points. 
 
Press Acquire  Mode in the function menu at the front panel and use 
  to 
select the desired acquisition mode (the default is normal), then press down the 
knob to select this mode. You can also press Mode continuously to switch the 
acquisition mode.   
 
 
Normal 
 
In this mode, the oscilloscope samples the signal at equal time interval to rebuild the 
waveform. For most of the waveforms, the best display effect can be obtained using 
this mode. 
 
 
Average 
 
In this mode, the oscilloscope averages the waveforms from multiple samples to 
reduce the random noise of the input signal and improve the vertical resolution. The 
greater the number of averages is, the lower the noise will be and the higher the 
vertical resolution will be but the slower the response of the displayed waveform to 
the waveform changes will be.   
 
The available range of the number of averages is from 2 to 8192 and the default is 2. 
When “Average” mode is selected, press Averages and use 
  to set the desired 
number of averages as the power function of 2. 
 
The Waveform before Average: