DS1000Z User’s Guide 
5-10 
Pulse Trigger 
 
Trigger on the positive or negative pulse with a specified width.   
 
Trigger Type: 
Press Type to select “Pulse”. At this point, the trigger setting information as shown in 
the figure below is displayed at the upper right corner of the screen.   
 
 
Source Selection: 
Press Source to select CH1-CH4 as th
. The current trigger source 
is displayed at the upper right corner of the screen. 
Note: select channel with signal input as trigger source to obtain stable trigger.   
 
Pulse Condition:   
Press When to select the desired pulse condition. 
 
: trigger when the positive pulse width of the input signal is greater than 
the specified pulse width. 
 
: trigger when the positive pulse width of the input signal is lower than the 
specified pulse width. 
 
: trigger when the positive pulse width of the input signal is greater than 
the specified lower limit of pulse width and lower than the upper limit of pulse 
width. 
 
: trigger when the negative pulse width of the input signal is greater than 
the specified pulse width. 
 
: trigger when the negative pulse width of the input signal is lower than 
the specified pulse width. 
 
: trigger when the negative pulse width of the input signal is greater than 
the specified lower limit of pulse width and lower than the upper limit of pulse 
width. 
 
Pulse Width Setting:   
In this oscilloscope, positive pulse width is defined as the time difference between 
the two crossing points of the trigger level and positive pulse as shown in the figure 
below.