MIL-STD-1553 Decode Setup Detail 
For general information about using controls shown on the main Serial Decode dialog, refer to Accessing The D 
and TD Supported Protocol Toolsets
 (on page 13)
MIL-STD-1553
 
B
ASIC AND 
L
EVELS 
R
IGHT
-H
AND 
D
IALOGS
 
Access the Serial Decode dialog by touching Analysis → Serial Decode on the menu bar. 
Touch the corresponding Setup... button for your decode. The Decode Setup... along with corresponding right-
hand dialogs are shown. 
MIL-STD-1553 Basic Right-Hand Dialog 
The Basic right-hand dialog provides detailed controls and setup conditions as follows: 
 
 
Table Mode - Choosing Word or Transfer shows/hides different columns from the table display. Word 
displays the Status column on your table display; Transfer shows RespDataIMG, and Status columns on 
your table display. 
 
Viewing Mode - Select to view the protocol data in Binary or Hexadecimal mode. 
 
Response Time - Provide From and To time lengths as desired. 
 
Inter Message Gap - Enter a Minimum threshold value for this parameter. 
MIL-STD-1553 Level Right-Hand Dialog 
The Level right-hand dialog provides detailed controls and setup conditions as follows: 
Adjust levels using Absolute or Percent Types. While High and Low values can be modified, preset levels are 
initially set to cross small amplitude signals. 
 
 
PLEASE NOTE THE FOLLOWING:  
 
Choose voltage or percentage level values carefully. High and Low values you provide are applied across 
all amplitudes. 
 
While MIL-1553 messages can contain varied amplitudes, they are still decoded. 
Take an example where a transaction contains a 20 V amplitude on the bus controller word and 2 V on 
the reply words of the remote terminal; a ratio of 2:10. 
The gain then has to be adjusted to decode the lowest amplitude words (2 V) with a gain of, say, 1 V/div. 
While the high amplitude words (20 V) are overflow, they are still decoded.