Lecroy WSXS-I2CBUS TD WSXS-I2CBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® oscilloscopes. WSXS-I2CBUS TD Benutzerhandbuch

Produktcode
WSXS-I2CBUS TD
Seite von 204
Operator's Manual 
919586 RevA 
47 
The Filter Right-Hand Dialog 
Use this dialog to specify which primitives for the protocol standard (selected from the basic Right-Hand Dialog, 
previously) are Decoded or Filtered out of the results. 
 
Using 8b/10b Hi-Speed Serial Triggers on Zi Oscilloscopes 
S
OMETIMES 
R
EFERRED TO AS 
S
ERIAL 
T
RIGGER
 
 WARNING 
Prevent serial trigger module damage and never apply external voltages to the clock and data output 
connectors.
 
SDA 7 Zi and 8 Zi model oscilloscopes include a serial pattern trigger and a Clock and Data Recovery module. 
PLEASE NOTE THE FOLLOWING:  
 
SDA 725 Zi and SDA 735 Zi model oscilloscopes include the WavePro Medium Speed Serial Trigger module 
option (WPZi-MSPT). The CDR range is 100 Mb/s - 1.25 Gb/s. 
 
SDA 740 Zi and SDA 760 Zi model oscilloscopes include the WavePro High Speed Serial Trigger module 
option (WPZi-HSPT). The CDR range 100 Mb/s - 2.7 Gb/s, 3.0, 3.125 Gb/s.  
 
 All of the SDA 8 Zi and Zi-A model oscilloscopes include the WaveMaster High Speed Serial Trigger 
module option (WM8Zi-HSPT). The CDR range 100 Mb/s - 2.7 Gb/s, 3.0, 3.125 Gb/s. 
The CDR module is built into the instrument and is accessible through the channel 4 input. The signal on channel 
4 is always present at the input to the serial pattern trigger module. The serial trigger also includes two outputs 
located on the front panel, to the right of the channel 4 input. These two SMA connectors allow access to the 
recovered clock and data signals from the serial trigger module for signals up to 2.7 Gb/s. The signals at these 
connectors are nominally zero mean with a peak-to-peak amplitude of 330 mV. 
Note: The recovered clock and recovered data output signals are not available for 3 and 3.125 Gb/s.