Lecroy WSXS-I2CBUS TD WSXS-I2CBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® oscilloscopes. WSXS-I2CBUS TD Benutzerhandbuch

Produktcode
WSXS-I2CBUS TD
Seite von 204
Serial Data Debug Solutions 
56 
919586 RevA 
Selection of Trigger Type results in dynamic changes to the I
2
Cbus Trigger dialog. Simple I
2
C triggers, such as 
Start, Stop, ReStart, and NoAck, require no additional setup, while frame-based triggers, such as ADDR, 
ADDR+DATA, FRAME LENGTH, and EEPROM require addition user-defined setup information. 
Select condition values by touching fields (using your finger, or use a mouse pointer). A pop-up is shown where 
you can choose from EqualNot EqualLess thanLess than or Equal toGreater thanGreater than or Equal to
In Range, or Out Range condition values. 
I
2
C Trigger Setup Detail 
The following topic provides specific control settings for an I
2
Cbus Trigger. 
 
The previously numbered I
2
Cbus trigger sections correspond with the following explanations. 
1.
 
S
OURCES 
S
ETUP
 
 
DATA and CLOCK - The pop-up dialog is used to select the appropriate channel or EXT inputs for each. Set 
these fields up with caution or your trigger may not function correctly. 
 
Threshold (Trigger) - Adjust the vertical level for the trigger. Much like an Edge trigger, a user must 
specify the level used in order to process the incoming signals and determine whether the desired serial 
data pattern is meeting the set trigger condition. This value is used for both DATA and CLOCK signals. 
2.
 
T
RIGGER 
T
YPE
 
The I
2
C trigger can be configured to trigger on simple conditions; meaning the presence of a STARTSTOP
RESTART bit, or the absence of an ACK bit (NO ACK). In addition, more complex trigger conditions can be 
created using ADDRADDR+DATAFRAME LENGTH, or EEPROM setups. 
If one of the more complex trigger conditions is selected, then reference the following sections for 
information on Address and Data Pattern Setup. 
3.
 
S
ETUP 
F
ORMAT
 
Select either Binary or Hexadecimal (Hex) setup mode. The format propagates through the entire I
2
trigger setup. 
A user can select Binary mode, and set up the address in binary format, then reselect Hex mode and set 
up the data in hexadecimal format. Toggling back and forth between the modes does not result in lost 
information (binary is used internally as the core format for all triggering and decoding operations), 
though use of don’t care bits in a binary setup results in the display of an X (for a full nibble don’t care) or 
a $ (for a partial nibble don’t care). 
4.
 
A
DDRESS 
S
ETUP
 
These following setup choices demonstrate ADDR, ADDR+DATA, FRAME LENGTH, or EEPROM Trigger 
Selections. 
 
Address Length - I
2
C utilizes either 7 or 10-bit formats for the address, depending on the device. Make the 
appropriate selection so as to be able to enter the correct address value.