Gw Instek GPT-9601 Insulation measuring device, GPT-9601 Benutzerhandbuch

Produktcode
GPT-9601
Seite von 21
 
 
GPT-9600 Series Quick Start Guide 
6
 
Cleaning the 
GPT-9600 
 
Disconnect the power cord before cleaning. 
 
Use a soft cloth dampened in a solution of mild 
detergent and water. Do not spray any liquid. 
 
Do not use chemicals containing harsh material 
such as benzene, toluene, xylene, and acetone. 
Operation 
Environment 
 
Location: Indoor, no direct sunlight, dust free, 
almost non-conductive pollution (Note below) 
 
Relative Humidity: ≤ 70% (no condensation) 
 
Altitude: < 2000m 
 
Temperature: 0˚C~40˚C 
 
(Pollution Degree) EN 61010-1:2010 specifies the pollution degrees 
and their requirements as follows. The GPT-9600 Series falls under 
degree 2. 
Pollution refers to “addition of foreign matter, solid, liquid, or 
gaseous (ionized gases), that may produce a reduction of dielectric 
strength or surface resistivity”. 
 
Pollution degree 1: No pollution or only dry, non-conductive 
pollution occurs. The pollution has no influence. 
 
Pollution degree 2: Normally only non-conductive pollution 
occurs. Occasionally, however, a temporary conductivity caused 
by condensation must be expected. 
 
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to 
condensation which is expected. In such conditions, equipment 
is normally protected against exposure to direct sunlight, 
precipitation, and full wind pressure, but neither temperature 
nor humidity is controlled. 
Storage 
environment 
 
Location: Indoor 
 
Temperature: -10°C to 70°C 
 
Relative Humidity: ≤ 85% (no condensation)