GPT-9600 Series User Manual 
22
 
 
6.  Only connect the test leads to the HIGH 
VOLTAGE terminals before the start of a test. 
Keep the test leads disconnected at all other 
times. 
 
7.  Always press the STOP button when pausing 
testing.  
 
8.  Do not leave the safety tester unattended. 
Always turn the power off when leaving the 
testing area.  
 
9.  When remotely controlling the safety tester, 
ensure adequate safety measures are in place to 
prevent: 
 
Inadvertent output of the test voltage. 
 
Accidental contact with the instrument during 
testing. Ensure that the instrument and DUT are 
fully isolated when the instrument is remotely 
controlled. 
 
10. Ensure an adequate discharge time for the 
DUT. 
When DCW or IR tests are performed, the DUT, 
test leads and probes become highly charged. 
The GPT-96XX has discharge circuitry to 
discharge the DUT after each test. The time 
required for a DUT to discharge depends on the 
DUT and test voltage. 
Never disconnect the safety tester before a 
discharge is completed.