Gw Instek GPT-9602 Insulation measuring device, GPT-9602 Benutzerhandbuch
Produktcode
GPT-9602
GPT-9600 Series User Manual
66
F
AQ
• The tester will not turn on.
• The panel keys are not working.
• When I press the START button the tester will not start testing.
• The accuracy does not match the specification.
• The panel keys are not working.
• When I press the START button the tester will not start testing.
• The accuracy does not match the specification.
The tester will not turn on.
Ensure the power cord is connected. Check to make sure the fuse is
not blown. See page 68.
not blown. See page 68.
The panel keys are not working.
Ensure the tester is not in SIGNAL I/O or Remote Connect mode,
page 56.
page 56.
When I press the START button the tester will not start
testing?
testing?
The tester must first be in the READY status before a test can be
started. Ensure the tester displays READY before pressing the
START button, page 42.
If “Double Action” is enabled, the START button must be pressed
within 0.5 seconds after the STOP button is pressed, otherwise the
tester will not start testing. See the Control Settings section on page
56.
started. Ensure the tester displays READY before pressing the
START button, page 42.
If “Double Action” is enabled, the START button must be pressed
within 0.5 seconds after the STOP button is pressed, otherwise the
tester will not start testing. See the Control Settings section on page
56.