OPERATION 
13 
 WARNING 
If Double Action is active, the START button must 
be pressed 500ms after the STOP button was 
pressed to be able to start testing. 
 WARNING 
If INTERLOCK is set to ON and the interlock 
connector is not connected to the SIGNAL I/O 
port, INTERLOCK OPEN will be displayed on the 
screen, preventing the test from starting. See page 
18 for the Common Utility settings. 
Return to VIEW Status 
 
Press the MANU/AUTO key when in READY status 
to return to VIEW status. 
 
MANU/AUTO
 
 
ZERO function (GB test only) 
 
Short the SOURCE H/L & SENSE H/L terminals using 
the GLT-115 alligator clips. 
 
In the READY status, press the ZERO soft-key. 
 
Press the START button to perform the zeroing. 
 
The tester will go into the ZERO status. 
 
When finished the tester will return to the VIEW status 
and the REF# will be set as measured resistance from 
the zeroing. 
E R O
Z
 
 
START
 
TEST Status 
 
Description 
TEST status is active when a MANU test or 
AUTO test is running. Pressing STOP will 
cancel the current test and any remaining steps 
in the case of an AUTO test. 