Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
 
GPT-9000 Series User Manual 
24
 
Operating Precautions 
 
Background 
The GPT-9000 is a high voltage instrument that 
outputs dangerous voltages. The following 
section describes precautions and procedures 
that must be followed to ensure that the tester 
is operated in a safe manner.  
 WARNING 
The GPT-9000 generates voltages of up to 5kVAC 
or 6kVDC. Follow all safety precautions, warnings 
and directions given in the following section when 
using the GPT-9000. 
 
1.  Never touch the safety tester, lead wires, 
terminals, probes and other connected 
equipment when the tester is testing. 
 
2.  Do not turn the safety tester on and off quickly 
or repeatedly. When turning the power off, 
please allow a few moments before turning the 
power back on. This will allow the protection 
circuits to properly initialize.  
Do not turn the power off when a test is 
running, unless in an emergency. 
 
3.  Only use those test leads supplied with the 
instrument. Leads with inappropriate gauges 
can be dangerous to both the operator and the 
instrument.  
For GB testing, never use the Sense leads on the 
SOURCE terminals.   
 
4.  Do not short the HIGH VOLTAGE terminal 
with ground. Doing so could charge the chassis 
to dangerously high voltages.