OPERATION
35
Manual Testing
This section describes how to create, edit and run a single ACW,
DCW, IR or GB safety test. Each Manual setting described in this
chapter only applies to the selected manual test – no other manual tests
are affected.
Each manual test can be stored/recalled to/from one of 100
memory locations. Each stored manual test can be used as a test
step when creating an AUTO test (page 79).
DCW, IR or GB safety test. Each Manual setting described in this
chapter only applies to the selected manual test – no other manual tests
are affected.
Each manual test can be stored/recalled to/from one of 100
memory locations. Each stored manual test can be used as a test
step when creating an AUTO test (page 79).
Choose/Recall a Manual Test number → from page 36.
Edit Manual Test Settings → from page 37.
Setting the Test Function→ from page 38.
Setting the Test Voltage or Test Current→ from page 38.
Setting the Test Frequency → from page 39.
Setting the Upper and Lower Limits → from page 40.
Setting a Reference Value → from page 42.
Setting the Test Time (Timer) → from page 43.
Setting the Ramp Up Time → from page 45.
Creating a MANU Test File Name→ from page 46.
Setting the ARC Mode → from page 47.
Setting PASS HOLD→ from page 49.
Setting FAIL HOLD → from page 50.
Setting MAX HOLD → from page 51.
Setting the Grounding Mode → from page 52.
Running a MANU Test → from page 58.
PASS / FAIL MANU Test → from page 62.
Zeroing of the Test Leads (GB only) → from page 67
Special MANU Test Mode (000) → from page 70
Before operating the GPT-9000 please read the safety precautions as
outlined in the Set Up chapter on page 20.