Gw Instek GPT-9801 Insulation measuring device, GPT-9801 Benutzerhandbuch

Produktcode
GPT-9801
Seite von 156
 
 
GPT-9000 Series User Manual 
48
 
 
3.  Use the scroll wheel to set the ARC 
mode. 
 
 
 
ARC MODES: 
OFF, ON AND CONTINUE,  
ON AND STOP 
 
4.  Press the EDIT/SAVE key to save 
and exit the MANU Utility and go 
back to EDIT status. 
EDIT/SAVE
 
 Note 
The ESC key can be pressed at any time in the 
Utility menu to cancel and exit. 
 
5.  If the ARC MODE was set to either ON AND 
CONTINUE, or ON AND STOP, the ARC 
current level can be edited. 
 
6.  Use the UP/DOWN arrow keys to 
move the cursor to the ARC setting. 
 
 
7.  Use the scroll wheel to edit the 
ARC level. 
 
 
 
ACW 
DCW 
1.000mA~080.0mA 
1.000mA~020.0mA 
 Note 
The ARC setting range is directly related to the HI 
SET current limit.  
 
 
ACW 
 
 
HI SET Limit 
ARC range 
 
 
0.001mA~0.999mA 
1.000mA ~2.000mA 
 
 
01.00mA~09.99mA 
01.00mA ~20.00mA 
 
 
010.0mA~042.0mA 
001.0mA ~080.0mA