52
 
 
2.  Use the UP/DOWN arrow keys to 
move to the MAX HOLD setting. 
 
 
3.  Use the scroll wheel to set MAX 
HOLD. 
 
 
 
MAX HOLD 
OFF, ON 
 
4.  Press the EDIT/SAVE key to save 
and exit the MANU Utility menu. 
EDIT/SAVE
 
 Note 
The ESC key can be pressed at any time in the 
MANU Utility menu to cancel and exit. 
Setting the Grounding Mode 
 
Background 
When GROUND MODE is set to ON, the GPT-
9000 grounds the return terminal to the ground. 
This mode is best for DUTs that are grounded 
to an earth ground by their chassis, fixtures or 
operation environment. This mode measures 
the potential of the HIGH VOLTAGE terminal 
with respect to earth ground. This means that 
any stray capacitance/resistance that leaks to 
earth ground will also be measured. This is the 
safest testing mode, though potentially not as 
accurate. 
When GROUND MODE is set to OFF, the 
return terminal is floating with respect to the 
earth ground. This mode is for DUTs that are 
floating and not directly connected to an earth 
ground. This is more accurate than when 
GROUND MODE is set to ON as any stray 
capacitance/resistance that leaks to the earth