GPT-9000 Series User Manual 
76
 
 
7.  Press EDIT/SAVE to save the 
settings and exit to VIEW status. 
EDIT/SAVE
 
 Note 
Ensure the baud rate settings or GPIB address 
matches the host machine. 
 Note 
The ESC key can be pressed at any time to cancel 
and exit back to VIEW status. 
Control Settings 
 
Description 
The Control settings are accessed in the 
COMMON UTILITY menu. The Control 
settings include: Start Control, Double Action, 
Key Lock and Interlock. 
Start Control is used to determine how a test is 
started. Tests can be started via the front panel 
(START/STOP buttons), from a remote 
controller or via the SIGNAL I/O port. 
The Double Action function is a safety feature 
used to prevent accidentally starting a test. 
Normally to start a test, the START button is 
pressed when the tester is in the READY status. 
To start a test when Double Action is ON, the 
STOP button must first be pressed, followed by 
the START button within 500ms.  
Key Lock disables the front panel keys from 
changing the test number, mode or testing 
parameters. Only the Utility menu and any 
keys required for testing are not disabled.