GPT-9000 Series User Manual 
98
 
Remote Controller Operation 
 
Description 
The GPT-9000 accepts external remote 
controllers with a START and STOP button. To 
use the REMOTE terminal, the GPT-9000 must 
first be configured to accept a remote controller.  
Operating a remote controller is the same as 
operating the START and STOP buttons on the 
front panel. 
Steps 
1.  Turn off the power. 
 
2.  Insert the lead of remote controller 
into the REMOTE terminal. 
REMOTE
  
 
3.  Turn on the power to the tester. 
 
 
4.  Configure the Start Ctrl option to 
REMOTE CONNECT in the 
Common Utility menu. 
Page 76 
 
5.  The tester will now only be able to 
start a test using a remote 
controller. 
 
 
6.  To return the operation control to 
the front panel, configure the Start 
Ctrl option to FRONT PANEL. 
Page 76