GPT-9000 Series User Manual
144
Sweep Commands
SWEEP:DATA:STATus ............................................................................ 144
SWEEP<X>:DATA:SHOW ...................................................................... 145
SWEEP:GRAPh:SHOW ............................................................................ 146
SWEEP :GRAPh:LINE .............................................................................. 146
SWEEP:STARt:TIME ................................................................................. 147
SWEEP:DATA:STATus
Query
Description
Returns the sweep mode, the voltage and current
settings and the number data points that are used
in the last sweep. There can be a maximum of 190
data points, depending on the testing time.
The data is returned as a string in the following
format:
SWEEP MODE,VSET,ISET,Get Data[#data points].
settings and the number data points that are used
in the last sweep. There can be a maximum of 190
data points, depending on the testing time.
The data is returned as a string in the following
format:
SWEEP MODE,VSET,ISET,Get Data[#data points].
Query Syntax
SWEEP:DATA:STATus?
Return parameter <string>
SWEEP MODE, VSET+unit, ISET+units,
Get Data=number of data points
Get Data=number of data points
Example
SWEEP:DATA:STATus?
>ACW,V=0.108kV,HI=10.96 mA ,Get Data=011
>ACW,V=0.108kV,HI=10.96 mA ,Get Data=011