If “Interlock” is enabled, the interlock key must be inserted into the 
signal I/O port on the rear before a test can be started. See page 108 
for details. 
Lastly, ensure that the Start Ctrl setting is correctly configured in 
the Common Utility menu. For example, to enable the START 
button to start a test, ensure that the Start Ctrl setting is set to 
FRONT PANEL. See page 100 for details. 
The accuracy does not match the specification. 
 
Make sure the tester is powered on for at least 30 minutes, within 
+15°C~+35°C. This is necessary to stabilize the unit to match the 
specification. 
For more information, contact your local dealer or GWInstek at 
www.gwinstek.com / marketing@goodwill.com.tw.