Test Errors 
The following error messages or messages may appear on the GPT 
screen when configuring or running tests.   
 
Error Messages 
Description 
TIME ERR 
For ACW tests.  
GPT-98XX: 
TIME ERR is displayed when HI SET ≥ 
30.00mA~40.00mA and if the RAMP  time 
and the TEST TIME setting is > 240 seconds. 
GPT-99XX: 
TIME ERR is displayed when HI SET ≥ 
80.00mA~100.0mA and if the RAMP  time 
and the TEST TIME setting is > 240 seconds. 
 
OVER 50W (GPT-98XX)  For DCW tests. OVER 50W is displayed if the 
HI SET setting multiplied by the Voltage setting 
is greater than 50W (GPT-98XX only). 
OVER 100W (GPT-99XX)  For DCW tests. OVER 100W is displayed if the 
HI SET setting multiplied by the Voltage setting 
is greater than 100W (GPT-99XX only). 
I ERR 
For ACW, DCW tests. Shown when the current 
is set too high. 
SHORT 
Voltage is too low. Indicates that the DUT 
could be shorted. 
V ERR 
For ACW, DCW tests. Voltage is too high. 
V = 0 
For GB tests. Voltage is equal to 0. Check to 
see that the SENSE H or SOURCE H is not 
open. 
R ERR 
For IR tests. The voltage is too high or 
resistance=0Ω. Check to see whether the DUT 
or test lead is shorting.  
 
For GB tests. The resistance is too high.