GPT-9000 Series User Manual 
44
 
 
 
ACW (LO) 
 
DCW (LO) 
 
IR (HI) 
 
GB (LO) 
0.000mA~041.9mA (GPT-98XX) 
0.000mA~109.9mA (GPT-99XX) 
0.000mA~010.9mA (GPT-98XX) 
0.000mA~020.9mA (GPT-99XX) 
0001MΩ~9999MΩ, ∞ (GPT-98XX) 
0.001GΩ~50.00GΩ, ∞ (GPT-99XX) 
000.0mΩ ~ 649.9mΩ 
 Note 
The LO SET setting is limited by the HI SET 
setting. The LO SET limit cannot be greater than 
the HI SET limit.  
When setting the current, be aware that a 
maximum of 200VA can be set for ACW and 50W 
for DCW (GPT-98XX) or 500VA and 100W, 
respectively for GPT-99XX. 
Setting a Reference Value 
 
Background 
The REF# acts as an offset. The REF# value is 
subtracted from the measured current (ACW, 
DCW) or measured resistance (IR, GB).  
Steps 
1.  Press the UP / DOWN arrow keys 
to bring the cursor to the  
REF# setting. 
 
 