Gw Instek GPT-9904 Insulation measuring device, GPT-9904 Benutzerhandbuch

Produktcode
GPT-9904
Seite von 171
 
 
GPT-9000 Series User Manual 
74
 
Special MANU Test Mode (000) 
 
Special Test 
Mode Overview 
When MANU number 000 is selected, the 
special test mode is activated. Under the special 
test mode, the voltage can be changed during a 
test, in real time (ACW, DCW only). The test 
function can also be changed when in READY 
or VIEW status, unlike under normal operation.  
Separate settings can be saved under the special 
test mode for each of the testing functions: 
ACW, DCW, IR and GB. This means a different 
ACW, DCW, IR and GB test setup can be saved 
for MANU number 000.  
Sweep Function 
Overview 
The GPT-9903 and 9904 also has access to the 
sweep mode function. The sweep function 
creates a graph of one of the ACW, DCW, IR or 
GB tests in the special manual mode. The graph 
will plot the output voltage, current or 
resistance versus time. After the test has been 
completed, the test current, voltage or 
resistance at any point in time can be viewed in 
the graph. 
Below is an example of the resultant sweep plot 
of a DCW test where a DC voltage is ramped 
up to a user-defined level until the HI SET 
current level has been tripped or the test time 
runs out.