EXTERNAL CONTROL 
101 
Using the SIGNAL I/O to Start/Stop Tests 
 
Background 
To use the SIGNAL I/O port the Start Ctrl 
settings have to be set to SIGNAL I/O in the 
Common Utility menu.  
Panel operation 
1.  Set the Start Ctrl option to SIGNAL 
I/O. 
 
2.  Turn off the power. 
 
 
3.  Connect the Input/Output signals 
to the SIGNAL I/O port. 
 
 
4.  Turn the tester back on. 
 
 
5.  To start the testing, short the 
INPUT_STOP and INPUT_COM 
line for a minimum of 1ms to put 
the tester into READY status. 
 
 
6.  To start the testing, short the 
INPUT_START and INPUT_COM 
lines for a minimum of 1ms. 
 
 
7.  To stop the testing, temporarily 
short the INPUT_STOP and 
INPUT_COM line again.