Gw Instek GPT-9803 Insulation measuring device, GPT-9803 Benutzerhandbuch

Produktcode
GPT-9803
Seite von 156
 
APPENDIX 
149 
GPT-9000 Specifications 
The specifications apply when the GPT-9000 is powered on for at 
least 30 minutes at 15˚C~35˚C.  
Specifications 
 
 
Environment 
Range 
Temperature 
Humidity 
Warranty  
15˚C ~ 35˚C 
≤70% (No 
condensation) 
Operation 
0˚C ~ 40˚C 
≤70% (No 
condensation) 
Storage 
-10˚C ~ 70˚C 
≤85% (No 
condensation) 
Installation Location 
Indoors at an amplitude of up to 2000m. 
 
AC Withstanding Voltage 
Output Voltage Range 
0.100kV~ 5.000kV  
Output Voltage Resolution  
2V 
Output Voltage Accuracy 
 (1% of setting +5V) with no load 
Maximum Rated Load(Table1)  200 VA (5kV/40mA) 
Maximum Rated Current 
40mA 
0.001mA ~ 10mA(0.1kV≤V≤0.5kV) 
0.001mA ~ 40mA(0.5kV<V≤5kV) 
Output Voltage Waveform 
Sine wave 
Frequency 
50 Hz / 60 Hz 
Voltage Regulation 
 1% +5V 
[Maximum rated load → no load] 
Voltmeter Accuracy 
 (1% of reading+ 5V) 
Current Measurement Range  0.001mA~040.0mA 
Current Best Resolution 
1uA 
0.001mA(0.001mA~0.999mA) 
0.01mA(01.00mA~09.99mA) 
0.1mA(010.0~040.0mA) 
Current Measurement 
Accuracy 
± (1.5% of rdg + 30 counts) when HI SET<1.00mA 
± (1.5% of rdg + 3 counts) when HI SET≥1.00mA 
Window Comparator Method  Yes 
ARC DETECT 
Yes 
Rise-time Control Function 
Yes 
RAMP (Ramp Time) 
0.1~999.9S 
Continued…