Gw Instek GPT-9803 Insulation measuring device, GPT-9803 Benutzerhandbuch

Produktcode
GPT-9803
Seite von 156
 
OPERATION 
45 
Setting the Ramp Up Time 
 
Background 
The Ramp Up time is the total time taken for 
the tester to reach the test voltage level. The 
Ramp Up time starts after the initial time 
(100ms) with a start voltage of 50 volts. The 
Ramp Up time can be set from 000.1 to 999.9 
seconds. The Ramp Up time is only applicable 
for ACW, DCW and IR tests. 
Discharge time 
Test V
time
Start V
Initial time 
(100ms)
TEST TIME
RAMP
  
Steps 
1.  Use the UP/DOWN arrow keys to 
bring the cursor to the RAMP  
setting. 
 
 
I R  
G B  
 
m A
A CW
D CW
I / L
H
O
I M E
T
R
 
 
E        
 
 
F R E Q =  
0 H z  
6
0
   
100
k
 
V
 
       
R
A C
 
=
F F
O
I
D
E
 
 
                         
M
T I
E
= 0 0 1 . 0 S
R
0 0 m A
E
R F
=
#
0 .
0
M
M A U
N
N A
_
1 . 0 0 m A
0
L
E
S
= 0
T
0 0 . 1 S
0
A M P
=
R
 
 
T
2
M A N U = * * * - 0 0
cursor
 
 
2.  Use the scroll wheel to set the 
RAMP  value. 
 
 
 
ACW 
DCW 
IR 
000.1s~999.9s  
000.1s~999.9s 
000.1s~999.9s 
 Note 
The discharge time and initial test time cannot be 
edited.