Control Functions
4.1.3 Quick Measurements
Measurements performed using a suitable fixed measuring range 
are executed more quickly than those which utilize automatic 
range selection. Quick measurement is made possible with the 
following two functions:
• Manual measuring range selection, i.e. selection of the measuring 
range with the best resolution (see chapter 4.1.2)
or
• With  the  DATA function (see chapter 4.4). In this way, the 
appropriate measuring range is selected automatically after the 
first measurement and the second measurement is executed 
more quickly.
The selected measuring range remains active for the subsequent 
series of measurements with these two functions.
4.2
Zero Offset / Relative Measurements
Zero balancing or a reference value for relative measurements can 
be stored to memory depending upon deviation from the zero 
point:
The relevant reference or correction value is deducted individually 
for the respective measuring function as an offset from all future 
measurements and remains in memory until deleted, or until the 
multimeter is switched off.
Zero balancing and reference value adjustment can be used for 
auto-ranging, as well as for manual measuring range selection.
Zero Balancing
➭ Plug the measuring cables into the instrument and connect the 
free ends to each other, except for capacitance measurement 
in which case the ends of the cables are not connected to 
each other.
➭ Briefly press the ZERO | ESC key. 
The instrument acknowledges zero balancing with an acoustic 
signal, and the “ZERO 
ΔREL” symbol appears at the LCD. The 
value measured at the moment the key is pressed serves as a 
reference value.
➭ Zero balancing can be cleared by once again pressing the 
ZERO | ESC key.
Note

As a result of TRMS measurement, the multimeter displays a 
residual value of 1 to 10/35 digits with short-circuited 
measurement cables as the zero point for V AC / I AC or 
V(AC+DC) / I (AC+DC) measurements (non-linearity of the 
TRMS converter). This has no influence on specified 
accuracy above 1% of the measuring range (or 3% in the 
mV, V(AC+DC) ranges).
Setting the Reference Value
➭ Plug the measuring cables into the instrument and measure a 
reference value (max. 1500 digits).
Deviation from zero 
– with short-circuited measurement cables for V, 
Ω, A
– with open input for capacitance unit of measure: F
Display
0 to 200 digits
ZERO 
ΔREL
> 200 to 1500 digits
ΔREL