Control Functions
Example
The voltage measuring range is set manually to 30 V. 
The first measured value is 5 V and is stored to memory because 
it is greater than 10% of the measuring range (= 3 V), and is thus 
reliable above the background noise level. As soon as the 
measured values drops to less than 10% of the measuring range, 
i.e. amounts to less than 3 V which corresponds to removal of the 
test probes from the measuring point, the instrument is ready to 
store a new value.
4.4.1 Saving Minimum and Maximum Values – MIN/MAX Function
Minimum and maximum measured values applied to the 
measuring instrument’s input after the Min/Max function has been 
activated can be “frozen” at the display. The most important use 
of this function is the determination of minimum and maximum 
values during long-term measured value observation.
The Min/Max function can be activated in all measuring functions.
The Min/Max function has no effect on the analog display, at 
which the current measured value continues to appear.
Apply the measured quantity to the instrument and set the 
measuring range with the MAN / AUTO key before activating the 
Min/Max function.
The Min/Max function is deactivated by pressing and holding the 
DATA/MIN/MAX key (approx. 1 second), when the measuring 
function is changed or when the instrument is switched off and 
back on again.
Note

As opposed to the DATA function, the Min/Max function can 
also be used for temperature measurement.
Function
Min/Max
Press 
DATA /
Min/Max
Min. and Max.
Measured Values
Response from Instrument
Display
Aco
us-
tic 
Sig-
nal
Measured 
Value 
Digital
Max.
Min.
1
Activate 
and save
2 x brief
Are
saved
Current 
measured 
value
Max 
and 
min
Twice
2
Save and 
display
Brief
Storage continues 
in background,
new min. and max. 
values are displayed.
Saved min. 
value
Min.
Once
Brief
Saved max. 
value
Max.
Once
Return to 
1
Brief
Same as 1,
stored values are not deleted
Same as 1
Same 
as 1
Once
Stop
Long
Are deleted
Current 
measured 
value
Is 
deleted
Twice