5.8.5 Alternating Current Measurement with A AC and Hz Clip-On 
Current Transformer
Current/Current Transformer Output
When a clip-on current transformer is connected to the 
multimeter (mA/A input), all current displays appear with the 
correct value in accordance with the selected transformation 
ratio. The only prerequisite is that the current transformer is 
equipped with at least one of the below listed transformation 
ratios, and that the ratio has been previously selected in the 
following menu (
Cl ip 
≠ 0FF).
Current Clip Setup Menu
 
1nFo 
Z
 ... 
Z
 
SET 
 rAtE 
V
 ... 
V
 
CLIP 
 
1 / 10/100/1000 / 0ff 
UV
 
Transf. Ratios
CL IP
DMM Measuring Ranges
Clip Types
30 mA
300 mA
3 A
1:1 
1mA / 1 mA
30.00 mA
300.0 mA
3.000 A
1:10 
1 mA / 10 mA
300 mA
3.000 A
30.00 A
1:100 
1 mA / 100 mA
3.000 A
30.00 A
300.0 A
1:1000 
1 mA / 1 A
30.00 A
300.0 A
3 000.0 A
WZ12A, WZ12D, 
WZ11A, Z3511, 
Z3512, Z3514
MEASURE
SETUP
FUNC
ENTER
FUNC
ENTER
FUNC
ENTER
COM mA A
V
Temp
Ω MΩ
ISO
~
R
x
Current
R
i
R
i
 
≤ 50 Ω
A~
Hz
A~
Hz
0
03.50
A
AC
TRMS
0
050.1
Hz
AC
FUNC
ENTER
1:1000
1:1000
MEASURE
SETUP
 ... 
Z
 
SET 
 ... 
V
 
CL IP
FUNC
ENTER
FUNC
ENTER
  
UV
 
0ff
Measuring Ranges:
Hz: 1 Hz . . . 31 kHz
mA / A: see table
3 ranges: 300 Hz, 3 kHz and 30 kHz