Gmw TG UNI 1AVDE-tester DIN VDE 0701/ 0702 61000 00200 Benutzerhandbuch

Produktcode
61000 00200
Seite von 44
42
15. Tecnical data
Protective conductor resistance:
Measuring range
0 ... 2 
Ω
Display range
0 ... 1.000 
Ω
Resolution
1 m
Ω
Error
± 5 % ± 10 digit
Measurement current
min. 200 mA DC (+/–)
Insulation resistance:
Measuring range
200 k
Ω ... 200 MΩ
Display range
0.200 ... 200 M
Ω
Resolution
0.001 k
Ω
Error
± 5 % ± 10 digit
Measurement voltage
min. 500 V DC
Substitute leakage current:
Measuring range
0 ... 20 mA
Display range
0 ... 20 mA
Resolution
0,001 mA
Error
± 5 % ± 10 digit
Measurement voltage
220 V AC
Protective conductor current, direct:
Measuring range
0 ... 5 mA
Display range
0 ... 5 mA
Resolution
0.001 mA
Error
± 5 % ± 10 digit
Touch current, direct:
Measuring range
0 ... 1 mA
Display range
0 ... 1 mA
Resolution
0.001 mA
Error
± 5% ± 10 digit
Protective conductor current, differential current:
Measuring range
0 ... 5 mA
Display range
0 ... 5 mA
Resolution
0.001 mA
Error
± 5 % ± 10 digit
Touch current, differential current:
Measuring range
0 ... 1 mA
Display range
0 ... 1 mA
Resolution
0.001 mA
Error
± 5% ± 10 digit
Operating manual for the TG uni 1 appliance tester
Technical data