                                                                                                                                                                                                     _____________________________________________________                          _______________________________________________________________________________
8 Device Configuration
Note:  Some menu items will be shown depending on the actual device configuration
(e.g. some items are disabled when the logger contains data).
Please note the hints by the menu items.
Enter Configuration:
Press for 2 seconds.
The main menu will be
shown
 or
  
Navigation:
choose menu branch (*1)
choose the parameter (*2)
 / 
edit the parameter values (*2)
Stores the settings, jumps back to the main menu
Stores the settings, leaves configuration
Menu(*1)
Parameters/Values (*2)
Meaning
p.r.t.
 / 
Read Logg: single value-logger read out
(only when data existing! Please refer to chapter 12)
12
Set Configuration: Generic Settings
°C:
All temperature values are in degrees Celsius
°F:
All temperature values are in degrees
Fahrenheit
*
0.1° / 0.01°:
Resolution 0.1° / 0.01°
Auto:
Resolution is selected automatically
*
9.1
E.751
characteristic curve according to EN60751
USEr
User sensor curve (Predefined to EN60751
values, changeable by software GMHKonfig)
*
9.2
-2.50...2.50°C /
-4.50...4.50°F
Zero correction
oFF:
Zero displacement inactive (=0.0°)
*
9.3
-2.000...2.000:
Scale correction [in %]
oFF:
Scale correction factor inactive (=0.000)
*
9.4
1...30:
Average filter (period in seconds)
oFF:
Average filter inactive
*
9.5
1...120
Power-off delay in minutes. Device will be
automatically switched off as soon as this time
has elapsed if no key is pressed or no
interface communication takes place
oFF
Power-off function inactive (continuous
operation)
oFF:
No output function
SEr:
Output is serial interface
dAC:
Output is analogue output