Gw Instek GDS-2304A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2304A Datenbogen
Produktcode
GDS-2304A
COMMAND DETAILS
49
:CHANnel<X>:PROBe:RATio
Set
Query
Description
Sets or returns the probe attenuation factor.
Syntax
:CHANnel<X>:PROBe:RATio { <NRf> | ?}
Related
Commands
Commands
:CHANnel<X>:PROBe:TYPe
Parameter
<X>
Channel 1, 2, 3, 4
<NRf> Probe attenuation factor.
Return parameter <NR3> Returns the probe factor.
Example
:CHANnel1:PROBe:RATio 1.00E+0
Sets the Channel 1 probe attenuation factor to 1x
Sets the Channel 1 probe attenuation factor to 1x
:CHANnel<X>:PROBe:TYPe
Set
Query
Description
Sets or returns the probe type (voltage/current).
Syntax
:CHANnel<X>:PROBe:TYPe { VOLTage | CURRent | ?}
Related
Commands
Commands
:CHANnel<X>:PROBe:RATio
Parameter
<X>
Channel 1, 2, 3, 4
VOLTage Voltage
CURRent Current
Return parameter Returns the probe type.
Example
:CHANnel1:PROBe:TYPe VOLTage
Sets the Channel 1 probe type to voltage.
Sets the Channel 1 probe type to voltage.
:CHANnel<X>:SCALe
Set
Query
Description
Sets or returns the vertical scale. The scale depends
on the probe attenuation factor.
Note the probe attenuation factor should be set
before the scale.
on the probe attenuation factor.
Note the probe attenuation factor should be set
before the scale.