58
 
:MATH:ADVanced:POSition  
Set
Query
 
 
Description 
Sets the vertical position of the advanced math 
result, expressed in unit/div.  
Syntax 
MATH:ADVanced:POSition { <NRf> | ? } 
Parameter 
<NRf> 
Vertical position: -12e+0 to +12e+0  
(12 units/division to +12 units/division.) 
Return parameter  <NR3> 
Returns the vertical position. 
Example 
:MATH:ADVanced:POSition 1.0e+0 
Sets the position as 1.00 unit/div. 
:MATH:ADVanced:SCALe 
Set
Query
 
 
Description 
Sets or queries the vertical scale the advanced 
math result.  
Syntax 
:MATH:ADVanced:SCALe {<NRf> | ?} 
Parameter 
<NRf> 
Vertical scale  
Return parameter  <NR3> 
Returns the vertical scale. 
Example 
:MATH:ADVanced:SCALe 2.0E-3 
 Sets the vertical scale to 2mV/S