GDS-2000A Series User Manual 
106
 
Show Waveform in XY Mode 
 
Background 
The XY mode maps the input of channel 1 to the 
input of channel 2. In 4 channel models, the input 
of channel 3 can be mapped to the input of channel 
4. This mode is useful for observing the phase 
relationship between waveforms. 
Reference waveforms can also be used in XY 
mode. Ref1 is mapped to Ref2 and Ref3 is mapped 
to Ref4. Using the reference waveforms is the same 
as using the channel input waveforms. 
Connection 
1.  Connect the signals to 
Channel 1 (X-axis) and 
Channel 2 (Y-axis) or 
Channel 3 (X2-axis) and 
Channel 4 (Y2-axis). 
CH3
CH2
CH1
CH4
Y
X
X2
Y2
 
 
2.  Make sure a channel pair is active 
(CH1&CH2 or CH3&CH4). Press 
the Channel key if necessary. A 
channel is active if the channel key 
is lit. 
CH1
CH1
 
CH2
CH2
 
Panel Operation 
1.  Press the Acquire menu key. 
Acquire
 
 
2.  Press XY from the bottom menu. 
 
 
3.  Choose Triggered XY from the side 
menu.