The calibration for Channel 1 starts and ends 
automatically, in less than 5 minutes. A 
message is displayed when the calibration 
procedure has ended. 
 
9.  Repeat the above step for Channel 2, 3* and 4* 
when prompted. 
*4 channel models only.
 
 
10. When the calibration for all channels has 
completed, the display goes back to the default 
state. 
Probe Compensation 
 
Panel Operation 
1.  Connect the probe between the Channel 1 input 
and the probe compensation output (Demo 3 
output, by default set as 2Vp-p, 1kHz square 
wave) on the front panel. Set the probe 
attenuation to x10. 
2.  Alternatively, the probe compensation signal 
can be changed. See page 171 for details. 
 
 