Calibration Screen
The data flash configuration can be saved to a file by selecting | File | Export | and entering a file name. A
data flash file also can be retrieved in this way, imported, and written to the bq3060 using the | Write All |
button.
The configuration information of the bq3060 and module calibration data also is held in the bq3060 data
flash.
The bq3060 allows for an automatic data flash export function, similar to the SBS Data logging function.
This feature, when selected via | Options | Auto Export |, exports Data Flash to a sequential series of files
named as FilenameNNNNN.gg where N = a decimal number from 0 to 9.
The AutoExport interval is set under the | Options menu | with a minimum value of 15 s. The AutoExport
filename also is set under the | Options menu |.
When a check mark is next to | AutoExport |, the AutoExport is in progress. The same menu selection is
used to turn on / off AutoExport.
If the data flash screen is blank, then the bq3060 that is being used may not be supported by the
bqEVSW version that is being used. An upgrade may be required.
9
Calibration Screen
9.1
How to Calibrate
Before the bq3060 is calibrated:
Connect a load to Pack- and Pack+ that draws approximately 2 A and measures discharge current to
use the FETs.
Connect a current source to Batt–(1N) and Pack- to calibrate without using the FETs.
Measure individual cell stack voltage from Batt–(1N), to Cell1(1P), Cell1+2(2P), Cell1+2+3(3P), and
Cell1+2+3+4(4P).
Measure the temperature of the pack.
These steps may not be required, depending on the type of calibration being performed.
Note that voltage calibration with cells attached requires special consideration. Cells must be in a resting
state. For additional information, go to the TI Web site (
and search for bq3060 Calibration
Using EV Software.
9.2
To Calibrate the bq3060
Select the types of calibration to be performed (see
).
Enter the measured values for the types selected.
If Software Voltage Calibration is selected, then enter the number of cells on the pack and individual cell
voltage.
If Temperature Calibration is selected, then select the sensor that is to be calibrated.
If the load is connected between Pack+ and Pack-, then select the Use FETs check box.
Press the Calibrate Part button.
9.3
Board Offset Calibration
This performs the offset calibration for the current offset of the board.
Remove load/external voltage and short Pack– to Batt–.
Press the CC Board Offset Calibration button.
13