Advanced bqMTester Instructions
Also, depending on items selected in the StationSetup Program, the Update VTI screen is populated
accordingly. As seen in
there is only one difference between the previous and following setups.
The following setup does not have Voltage selected in the StationSetup Program; thus, the voltage
controls are not populated during calibration.
Figure 12. Update VTI: No Voltage Selected
4.7.2
Reference Adjustments
Once unlocked, the references can be adjusted as required. When any field is selected in a particular
station row, the LEDs for that station flash and the voltage and current power sources power up.
To calibrate the references, use the following process:
1. Measure the voltage for the first station by connecting a traceable DMM to the Reference V Meter +
and Reference V Meter
connections (as shown as DMM1 in
) to measure the actual voltage
of the cell simulation voltage supplied by the calibration board for the first station. Input this voltage in
the Voltage column for the first station. Repeat this step for each remaining station.
2. Setup the DMM for current measuring, and connect the DMM to Reference I Meter + and Reference I
Meter
(as shown as DMM2 in
) for the first station being setup. Be sure to disconnect the
wire that shorted these two connections so that current flows through the meter. Input the current
measured in the Current column for this station. Repeat this step for each remaining station. Re-install
the short from the Reference I Meter + to the Reference I Meter
.
3. Place the traceable temperature probe next to the temperature probe being used on the calibration
board at the first station with a temperature probe being used for testing. Click Read the Currently
Calibrated Temperatures
. Compare the temperature from the traceable temperature probe to the
calibration board temperature displayed. If the temperatures are different, then type in the temperature
from the traceable temperature probe into the corresponding temperature field. Type over the value
displayed when Update VTI is selected. Repeat this step for each station that has a temperature
probe.
4. In the case of the bq30xx parts, additional voltage controls are populated depending on the number of
cells selected. The same procedure of calibration used in Step 1 can be used for the individual voltage
controls.
4.7.3
Allow V, T, I While Locked Selection
If the software is unlocked, then the Allow V, T, I while locked checkbox is enabled. Otherwise, it is
disabled and dimmed. If the checkbox is selected, you can adjust the actual values for voltage,
temperature, and current references even though the configuration has been locked. If the checkbox is not
selected, you cannot alter these values without unlocking the configuration.