Texas Instruments LMP90100 Sensor AFE Evaluation Board LMP90100EB/NOPB LMP90100EB/NOPB Datenbogen

Produktcode
LMP90100EB/NOPB
Seite von 46
 
 
 
March 2009 
 
 
 
 
 
 
  
 
SNAU028A 
21 
 
6.0.  Example #4: Thermocouple and LM94022 Application 
 
A.  Thermocouple and Cold Junction Compensation Background 
 
 
Figure 19 – Thermocouple and Temperature Sensor Connection 
 
As described in section 17.6.2. of the LMP90100 datasheet, because a thermocouple can only 
measure a voltage difference and thus a temperature difference (relative temperature), it does not 
have the ability to measure absolute temperature.  To determine the absolute temperature of the 
measured environment, a technique known as cold junction compensation (CJC) must be used.   
 
In a CJC technique, the “cold” junction temperature, Tcold (Figure 19), is sensed by using an IC 
temperature sensor, such as the LM94022.  The temperature sensor should be placed within close 
proximity of the reference junction.  The LM94022 is placed underneath the thermocouple 
connector J4 on the LMP90100 evaluation board.   
 
The technique to calculate for Thot using the CJC method can be found in the LMP90100 
datasheet.   
 
The Sensor AFE software does have the ability to display the relative thermocouple temperature 
using a Type K look-up-table (
).  In 
addition, if the LM94022 (or any other temperature sensors) is connected to the LMP90100, then 
the software can also read its temperature (Tcold).  However, the user has to manually enter this 
Tcold value in the 
 field located on the upper 
left hand side of the GUI.  The software will use this T_board value to calculate for Thot.  
B.  Thermocouple and LM94022 Schematic on the LMP90100EB