Texas Instruments LMP90100 Sensor AFE Evaluation Board LMP90100EB/NOPB LMP90100EB/NOPB Datenbogen

Produktcode
LMP90100EB/NOPB
Seite von 46
 
 
 
28 
LMP90100 EVB User’s Guide   
 
 
 
 
 
            
 
   August 2012 
Figure 25 – Temperature Sensor Reading for Tcold 
 
6.  Enter the LM94022’s mean temperature (see red box in the figure above) in the 
T_board” field located on the upper left hand side of the GUI.  The software will use 
this “T_board” value to calculate for the thermocouple’s absolute temperature (Thot). 
 
If the LM94022 or any other temperature sensor is not connected to do cold junction 
compensation, then the user can still manually enter a Tcold value in the “T_board” field 
box. 
 
7.  In the Sensor Window, click on CH1 to read the thermocouple’s voltage and temperature. 
 
Figure 26 – Sensor Window 
 
8.  In the “Measurement” tab, choose the “Output Format” as “Output Voltage (V)”
and click “Run” to capture the thermocouple relative voltage.  
 
 
Figure 27 – Thermocouple Voltage Relative Reading