Running the Software
5. No. of samples = 512
Data rate (sps) = 500
Show data for the last 2 secs
# of samples for FFT which is power of 2 = 512
min ( (500 x 2) , 512 )
Since (# of samples for FFT calc. == No. of samples) and If (Filter Type = None) then # of
samples for FFT calc. which is power of 2 = 512
Since (# of samples for FFT calc. == No. of samples) and If (Filter Type = Notch) then an error
msg “Insufficient # of samples for FFT calculation” will be displayed since # of samples for FFT
calc. which is power of 2 = (512 / 2) < 512 samples
Figure 25. ADC Capture and Analysis Tab
Figure 26. Scope Analysis: Test Results
22
AFE4400 and AFE4490 Development Guide
SLAU480A – January 2013 – Revised February 2013
Copyright © 2013, Texas Instruments Incorporated