Ea Elektro Automatik EA Elektro-Automatik EA-PS 2042-20B 320W 1 Output Programmable DC Power Supply, Switched Mode, Bench 39200114 Benutzerhandbuch

Produktcode
39200114
Seite von 20
18
EN
Instruction manual
PS 2000 B Single Series
Date: 27-02-2012
PS 2084-03B
PS 2084-05B
PS 2084-10B
Mains input
Input voltage
100…240V ±10%
100…240V ±10%
100…240V ±10%
Frequency
45…65Hz
45…65Hz
45…65Hz
Fuse
T 2A
T 3.15A
T 6.3A
Power factor
> 0.99
> 0.99
> 0.99
Power consumption at output off
12W
12W
15W
Output - Voltage
Adjustable range
0...84V
0...84V
0...84V
Stability at mains fluctuation
±10% ∆U
IN
< 0.02%
< 0.02%
< 0.02%
Stability at 0…100% load
< 0.15%
< 0.15%
< 0.15%
Ripple HF @ BWL 20MHz
< 60mV
PP
< 60mV
PP
< 150mV
PP
Ripple LF @ BWL 300kHz
< 10mV
RMS
< 10mV
RMS
< 2.1mV
RMS
Accuracy*
≤ 0.2%
≤ 0.2%
≤ 0.2%
Overvoltage protection
0…92.4V
0…92.4V
0…92.4V
Regulation time 10-90% load
< 2ms
< 2ms
< 2ms
Rise time 10-90% U
max
~165ms
~165ms
~165ms
Output - Current
Adjustable range
0…3A
0…5A
0…10A
Stability at mains fluctuation
±10% ∆U
IN
< 0.05%
< 0.05%
< 0.05%
Stability at 0…100% ∆U
OUT
< 0.15%
< 0.15%
< 0.15%
Ripple HF @ BWL 20MHz
< 6mA
PP
< 9mA
PP
< 3.8mA
PP
Ripple LF @ BWL 300kHz
< 2mA
RMS
< 3mA
RMS
< 1.3mA
RMS
Accuracy*
≤ 0.2%
≤ 0.2%
≤ 0.2%
Output - Power
Efficiency
≤ 85%
≤ 85%
≤ 85%
Nominal power P
nom
100W
160W
320W
Miscellaneous
Operation temperature
0….50°C
0….50°C
0….50°C
Storage temperature
 -20….70°C
 -20….70°C
 -20….70°C
Humidity rel.
< 80%
< 80%
< 80%
Dimensions of enclosure (WxHxD)
174x82x240mm
174x82x240mm
174x82x267mm
Dimensions absolutely (WxHxD)
174x90x263mm
174x90x263mm
174x90x290mm
Weight
1.9kg
2kg
2.6kg
Cooling
fanless,
natural convection
fanless,
natural convection
temperature
controlled fan
Noise
-
-
<40dBA @ 1m
Safety
EMC standards
Overvoltage class
Protection class
Article number
39200116
39200117
39200118
EN 60950
EN 61326, EN 55022 Class B 
Class II
Class I
* Related to the nominal value, the accuracy defines the maximum allowed deviation between set value and actual value.
Example: a 42V model has min. 0.2% voltage accuracy, this is 84mV. When setting a voltage of 5V and with an allowed 
maximum deviation of 84mV, the resulting actual value could be between 4.92V and 5.08V. 
Operating the device
8.2  84V models