Model 6485 and 6487 User’s Manual
Measurements and Sourcing Voltage
3-13
Sourcing voltage 
`^rqflk
Do not connect external sources to the Model 6487 voltage source. Exter-
nal sources may damage the Model 6487 voltage source.
Turning source on and off
The OPER key will toggle the voltage source state (on/off). The Model 6487 will output 
the configured voltage and the VOLTAGE SOURCE OPERATE indicator will turn on. 
OPER (operate) key
The OPER (operate) key will function to turn the voltage source off, even if the instrument 
is operating under remote control (REM annunciator on), assuming that the LLO (Local 
Lockout) function has not been employed. While in remote, the OPER key will only turn 
the source off. To turn it on, the Model 6487 must be in local (see 
).
Voltage source off state
The voltage source is not in a high-impedance state when it is turned off. Rather, it is in a 
state that acts just like the voltage source was programmed to 0V on the selected range. It 
will enter this state on power-up after the VOLTAGE SOURCE OPERATE light blinks. 
In contrast, the safety interlock will cause the voltage source to go into a high-impedance 
state instead of 0V output and the source will stay in the high-impedance state until the 
operate state is changed to on. The exception is the 10V range where the interlock is 
optional. The OPERATE light and front panel display do not indicate the difference 
between 0V output and high-impedance output caused by an open interlock. The interlock 
status is available by query via remote (see 
Compliance indication
At any time, it is possible that the voltage source will go into compliance (current limit 
reached). Should this situation occur, the OCOMP annunciator (Output Compliance) will 
flash and the displayed voltage value (visible when current readings are configured for 
readings of less than 6  digits) will alternate between showing the value and displaying 
“CMPL”. If you are in a menu where the voltage source value is not shown on the right-
most four characters of the display, only the flashing OCOMP annunciator will be shown.
Table 3-6
Voltage source current limits
Source Range:
Selectable Current Limit
10.0000V Range
25µA
 250µA 
 2.5mA 
 25mA
50.000V Range
25µA
 250µA 
 2.5mA
500.00V Range
25µA
 250µA 
 2.5mA