Model 6485 and 6487 User’s Manual
Range, Units, Digits, Rate, and Filters
4-3
Units
Readings can be displayed using engineering (ENG) units (i.e., 1.236 m) or scientific 
(SCI) notation (i.e., 1.236E-03). To change units, press the MENU key, select UNITS, 
then press ENTER. Select ENG or SCI, then press ENTER. 
klqb
The units setting can only be changed from the front panel (no remote 
operation). Scientific notation provides more resolution on small values than 
engineering units.
Digits
The DIGITS key sets display resolution for Model 6485/6487. Display resolution can be set 
from 3  to 6  digits. This single global setting affects display resolution for all measurement 
ranges. To set display resolution, press (and release) the DIGITS key until the desired num-
ber of digits is displayed. 
klqb
For the Model 6487, the voltage source value will not be displayed in the 6  digit 
display mode.
SCPI programming — range and digits
Programming example — range and digits
The following command sequence selects the 20mA range and sets display resolution to 3  
digits.
*RST
' Restore RST defaults.
RANG 0.02
' Set to 20mA range.
DISP:DIG  4
' Set display resolution to 3-1/2 digits.
Table 4-2
SCPI commands — range and digits
Commands 
Description
RANG <n>
RANG:AUTO <b> 
RANG:AUTO:ULIM <n>
RANG:AUTO:LLIM <n>
DISP:DIG <n>
Select range: -0.021 to 0.021 (A). 
Enable (ON) or disable (OFF) autorange.
Specify upper range limit for autorange: -0.021 to 0.021 (A).
Specify lower range limit for autorange: -0.021 to 0.021 (A).
Set display resolution: 4 (3  digits) to 7 (6  digits).