Using the Current Monitor Output 
 
Background 
The current monitor is used to measure the 
voltage drop across the shunt resistors 
manually.  
The current monitor outputs the full scale 
current input (for the selected range) as a 
voltage of 0~300mV.  
Shunt Values 
 Range 
Shunt 
 
 
30 mA 
10Ω  
 
 
300 mA 
1Ω  
 
 
3 A 
0.1Ω  
 
 
30 A 
0.01Ω  
 
 
300 A 
0.001Ω  
Steps 
1.  Set the PCS-1000 for normal operation, as 
described previously in this chapter, page 
25~27.  
Make note of the range used and the shunt that 
is used for that range.  
 
2.  Connect the current monitor output to a DVM. 
3.  Use OHM’s law, V=IR, to determine the 
current across the shunt resistor.  
For example: 
If we are using the 3A current range (and thus 
the 0.1Ω shunt) and the current monitor 
outputs 150mV, then: 
Input current = monitor output / shunt Ω 
= 150mV/0.1Ω  
= 1.5A