PCS-1000 User Manual 
44
 
Interface Configuration 
Configure GPIB Interface  
To use GPIB the GPIB address must first be set. 
 
Configure GPIB 
1.  Connect the GPIB cable from the GPIB 
controller to the PCS-1000. 
 
2.  Turn the PCS-1000 on. 
 
3.  Long push Func key to enter the 
function menu.  
Page 31   
 
4.  Use the ◄ Func ► keys to go to the ADDRESS 
function. 
5.  Select the address using the ▼Select ▲ keys. 
 
GPIB Address 
00~30 
 
6.  Press the Enter key to confirm the selection. 
 Note 
RMT will be displayed on the screen when the unit 
is remote mode. 
GPIB constraints 
 
Maximum 14 devices altogether, 20m cable 
length, 2m between each device 
 
Unique address assigned to each device 
 
At least 2/3 of the devices turned On 
 
No loop or parallel connection