PCS-1000 User Manual
44
Interface Configuration
Configure GPIB Interface
To use GPIB the GPIB address must first be set.
Configure GPIB
1. Connect the GPIB cable from the GPIB
controller to the PCS-1000.
2. Turn the PCS-1000 on.
3. Long push Func key to enter the
function menu.
Page 31
4. Use the ◄ Func ► keys to go to the ADDRESS
function.
5. Select the address using the ▼Select ▲ keys.
GPIB Address
00~30
6. Press the Enter key to confirm the selection.
Note
RMT will be displayed on the screen when the unit
is remote mode.
is remote mode.
GPIB constraints
Maximum 14 devices altogether, 20m cable
length, 2m between each device
length, 2m between each device
Unique address assigned to each device
At least 2/3 of the devices turned On
No loop or parallel connection