The signal oriented measurement commands are used to acquire readings.  You can use 
these high-level instructions to control the measurement process. These commands are 
summarized in Table 5-1.
:CONFigure:<function> 
<function> = CURRent:AC
AC current
CURRent[:DC]
DC current
VOLTage:AC
AC voltage
VOLTage[:DC]
DC voltage
RESistance
2-wire resistance
FRESistance
4-wire resistance
PERiod
Period
FREQuency
Frequency
TEMPerature
Temperature
DIODe
Diode testing
CONTinuity
Continuity test
Query
:CONFigure?     Query the selected function. 
Description
This command configures the instrument for subsequent measurements 
on the specified function.  Basically, this command places the instrument 
in a “one-shot” measurement mode.  You then use the :READ? command 
to trigger a measurement and acquire a reading (see :READ?). 
When this command is sent, the Model 2000 will be configured as follows: 
• The function specified by this command is selected. 
• All controls related to the selected function are defaulted to the *RST 
values. 
• Continuous initiation is disabled (:INITiate:CONTinuous OFF). 
• The control source of the Trigger Model is set to Immediate. 
• The count values of the Trigger Model are set to one. 
Table 5-1
Signal oriented measurement command summary
Command
Description
:CONFigure:<function>
:FETCh?
:READ?
MEASure[:<function>]?
 Places the Model 2000 in a “one-shot” measurement mode for the  
specified function. 
Requests the latest reading without triggering. 
Performs an :ABORt, :INITiate, and a :FETCh?. 
Performs an :ABORt, :CONFigure:<function>, and a :READ?. 
2000-900 (J - Aug 2010) BOOK.fm  Page 3  Wednesday, October 12, 2011  12:29 PM