Keithley 2000/E Digital-Multimeter, DMM, 2000/E Datenbogen

Produktcode
2000/E
Seite von 256
To achieve a stable reading, it helps to shield resistances greater than 100k
Ω. Place the 
resistance in a shielded enclosure and connect the shield to the INPUT LO terminal of the 
instrument electrically.
See Section 3—Measurement Options for information that explains the configuration op-
tions for 2-wire and 4-wire resistance measurements.
2001 MULTIMETER
Model 2000
Resistance
Under Test
Shielded
Cable
Optional shield
Note:  Source current flows from the INPUT
          HI to INPUT LO terminals.