MEASure2:CURRent:DC? 
Returns the DC current measurement on the second display. 
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)] 
Example: MEAS2:CURR:DC? 
>+0.234E-4 
Returns the DC current measurement as 0.0234 mA. 
MEASure2:CURRent:AC? 
Returns the AC current measurement on the second display. 
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)] 
Example: MEAS2:CURR:AC? 
> +0.387E-2 
Returns the AC current measurement. 
MEASure2:RESistance? 
Returns the 2W resistance measurement on the second display. 
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)] 
Example: MEAS2:RES? 
> +1.1912E+3 
Returns the 2W measurement. 
MEASure2:FREQuency? 
Returns the frequency measurement on the second display. 
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)] 
Example: MEAS2:FREQ? 
> +2.3712E+2 
Returns the frequency (237.12Hz). 
MEASure2:PERiod? 
Returns the period measurement on the second display. 
Parameter: [None] | [Range(<NRf> | MIN | MAX | DEF)] 
Example: MEAS2:PER? MAX 
Returns the period at the maximum range.